Specifications
Name: | Veraview X800 |
Model: | X800 |
Order Selection: | F40 / R100 / F150 |
Rating: | AC 120V 60Hz AC 220 / 230 / 240V 50 / 60Hz |
Power Consumption: | 2.0 kVA |
Weight: | Approx. 185 kg (approx. 220 kg with cephalometric) |
Manufacturer: | J. MORITA MFG. CORP. |
X-ray Tube Voltage: | 60 – 100 kV (depending on exposure mode) |
X-ray Tube Current: | 2 – 10 mA (depending on exposure mode) |
Nominal Focal Spot: | 0.5 |
Panoramic Exposures: | High speed mode (standard panoramic) approx. 7.4 sec. Fine Mode (standard panoramic) approx. 14.8 sec. |
Panoramic Regions: (consistent magnification) |
Standard Panoramic (Standard, Orthographic, Jaw), Pedodontic Panoramic (Standard, Orthographic, Jaw), Maxillary Sinus Panoramic (Anterior, Posterior), Quadruple TMJ, Partial Panoramic, Bite-wing Exposure Distances measured on a panoramic image are not equal to the actual distances. |
CBCT Exposure Time: | Approx. 9.4 sec. (180º) / approx. 17.9 sec. (360º) |
CBCT Exposure Regions: | F40P / F40CP / F40C - Ø40 x H40, Ø40 x H80 |
R100P / R100CP / R100C - Ø40 x H40, Ø40 x H80 - Ø80 x H40, Ø80 x H50, Ø80 x H80 - R100 x H40, R100 x H50, R100 x H80 |
|
F150P / F150CP / F150C - Ø40 x H40, Ø40 x H80 - Ø80 x H40, Ø80 x H50, Ø80 x H80 - R100 x H40, R100 x H50, R100 x H80 - Ø150 x H50, Ø150 x H75, Ø150 x H140 |
|
Cephalometric | |
Exposures: | F40CP / F100CP |
Direction and Size: | LA 220 x 250, PA 220 x 200 mm |
Wear protective aprons and coverings as necessary during X-ray exposure. The unit must be fixed to the floor and wall when installed. If minimal layout dimensions are used, there may be very little space to move around inside the booth. |